Ion Spectra and Anisotropy (ION)

ResourceID
spase://CNES/Instrument/CDPP-Archive/Interball-Auroral/ION

Description

The ION experiment performs simultaneous ion and electron measurements. Two mass spectrometers looking in opposing directions perpendicular to the satellite spin axis, which points toward the sun, measure ions in the mass and energy ranges 1-32 amu and ~ 0 to 14,000 eV. The ion measurements are complemented by simple electron spectrometers performing measurements in the energy range ~10 eV (or 60 eV) to 20,000 eV.

Ion composition analyzers

The design of the detector heads of the ION-1 and ION-2 mass spectrometers is based on the principle of the Wien filter, where the mass filtering is performed by crossed electric and magnetic fields. The energy over charge (E/Q) selection is made by a cylindrical electrostatic analyzer placed in front of the filter. At each energy step, the electric field in the magnetic analyzer is programmed in order either to scan 2 or 4 ions (H+, He++, He+, O+).

Electron analyzers

Two spectrometers, ELEC-1 and ELEC-2 (also called ION-3 and ION-4), measure electron fluxes. They consist of two cylindrical electrostatic analyzer heads and associated electronics.

View XML | View JSON | Edit

Details

Version:2.4.0

Instrument

ResourceID
spase://CNES/Instrument/CDPP-Archive/Interball-Auroral/ION
ResourceHeader
ResourceName
Ion Spectra and Anisotropy (ION)
ReleaseDate
2022-04-26 08:14:26Z
Description

The ION experiment performs simultaneous ion and electron measurements. Two mass spectrometers looking in opposing directions perpendicular to the satellite spin axis, which points toward the sun, measure ions in the mass and energy ranges 1-32 amu and ~ 0 to 14,000 eV. The ion measurements are complemented by simple electron spectrometers performing measurements in the energy range ~10 eV (or 60 eV) to 20,000 eV.

Ion composition analyzers

The design of the detector heads of the ION-1 and ION-2 mass spectrometers is based on the principle of the Wien filter, where the mass filtering is performed by crossed electric and magnetic fields. The energy over charge (E/Q) selection is made by a cylindrical electrostatic analyzer placed in front of the filter. At each energy step, the electric field in the magnetic analyzer is programmed in order either to scan 2 or 4 ions (H+, He++, He+, O+).

Electron analyzers

Two spectrometers, ELEC-1 and ELEC-2 (also called ION-3 and ION-4), measure electron fluxes. They consist of two cylindrical electrostatic analyzer heads and associated electronics.