Data Access
This dataset contains operational parameters for the instrument (corresponding to a TC sent from ground). Change in operational parameters can have an impact on data time resolution, sensitivity and can influence the instrument response.
Version:2.4.0
This dataset contains operational parameters for the instrument (corresponding to a TC sent from ground). Change in operational parameters can have an impact on data time resolution, sensitivity and can influence the instrument response.
Role | Person | StartDate | StopDate | Note | |
---|---|---|---|---|---|
1. | PrincipalInvestigator | spase://CNES/Person/CDPP-Archive/Pierre.Henri |
Access via CNES Data Archive for CDPP
Interference frequencies to be suppressed during the extremum computation in active modes
Interference frequencies to be suppressed during the extremum computation in active modes
Interference frequencies to be suppressed during the extremum computation in active modes
Takes one of the following values: 1.000 full level (156 mV RMS in MIP mode, 30 V RMS in LDL mode) - 0.500 nominal level (78 mV RMS in MIP mode, 15 V RMS in LDL mode) - 0.250 low level (39 mV RMS in MIP mode, 7.5 V RMS in LDL mode) - 0.125 minimum level (19.5 mV RMS in MIP mode, 3.75 V RMS in LDL mode)
Takes one of the following values: mono with E1 - mono with E2 - stereo with E1-E2 phased - stereo with E1-E2 anti-phased
Takes one of the following values: mono with E1 - mono with E2 - stereo with E1-E2 phased - stereo with E1-E2 anti-phased
Possible values are MIP alone or LDL mode
Autoloop : an auto-loop can be set between emission and reception - The field takes one of the following values: MIP with sensor - MIP auto-loop ON
Coding level for the power spectrum in the Passive mode - The field takes one of the following values: 2 (16 values from 0 to 30 dB) - 4 (16 values from 0 to 60 dB) - with 0 dB = 0.6 microVHz*-0.5
Amplitude threshold for the extremum detection - The field takes one of the following values: 1 - 2 - 4 - 8 - with 0 dB = 0.6 microVHz*-0.5
Selection of the frequency bandwidth in Survey modes. Three parameters (Frequency bandwidth, Frequency resolution and Number of frequency steps) are given for all intervals (nominal frequency interval -0- and the 7 complementary ones - nr 1 to nr 7 - ) : 0 nominal interval from 28 to 3472 kHz 28 - 224 kHz 7 kHz 29 / 238 - 448 kHz 14 kHz 16 / 476 - 896 kHz 28 kHz 16 / 952 - 1792 kHz 56 kHz 16 / 1904 - 3472 kHz 112 kHz 15 - 1 interval nr 1 from 28 to 665 kHz 28 - 665 kHz 7 kHz 92 - 2 interval nr 2 from 259 to 896 kHz 259 - 896 kHz 7 kHz 92 - 3 interval n r3 from 518 to 1792 kHz 518 - 1792 kHz 14 kHz 92 - 4 interval nr 4 from 924 to 3472 kHz 924 - 3472 kHz 28 kHz 92 - 5 interval n r5 from 357 to 987 kHz 28 - 343 kHz 7 kHz 46 / 357 - 987 kHz 14 kHz 46 - 6 interval nr 6 from 28 to 1582 kHz 28 - 224 kHz 7 kHz 29 / 238 - 630 kHz 14 kHz 29 / 658 - 1582 kHz 28 kHz 34 - 7 interval nr 7 from 268 to 2184 kHz 266 - 896 kHz 14 kHz 46 / 924 - 2184 kHz 28 kHz 46
Selection of the frequency bandwidth in Sweep modes. Three parameters (Frequency bandwidth, Frequency resolution and Number of frequency steps) are given for all intervals (nominal frequency interval -0- and the 7 complementary ones - nr 1 to nr 7 - ) : 0 nominal interval from 28 to 3472 kHz 28 - 224 kHz 7 kHz 29 / 238 - 448 kHz 14 kHz 16 / 476 - 896 kHz 28 kHz 16 / 952 - 1792 kHz 56 kHz 16 / 1904 - 3472 kHz 112 kHz 15 - 1 interval nr 1 from 28 to 665 kHz 28 - 665 kHz 7 kHz 92 - 2 interval nr 2 from 259 to 896 kHz 259 - 896 kHz 7 kHz 92 - 3 interval n r3 from 518 to 1792 kHz 518 - 1792 kHz 14 kHz 92 - 4 interval nr 4 from 924 to 3472 kHz 924 - 3472 kHz 28 kHz 92 - 5 interval n r5 from 357 to 987 kHz 28 - 343 kHz 7 kHz 46 / 357 - 987 kHz 14 kHz 46 - 6 interval nr 6 from 28 to 1582 kHz 28 - 224 kHz 7 kHz 29 / 238 - 630 kHz 14 kHz 29 / 658 - 1582 kHz 28 kHz 34 - 7 interval nr 7 from 268 to 2184 kHz 266 - 896 kHz 14 kHz 46 / 924 - 2184 kHz 28 kHz 46
Possibility to inhibit the MIP watchdog . The field takes one of the following values: watchdog on - watchdog off
Selection of the telemetry rate.The field takes one of the following values: minimum rate - normal rate - burst rate
Selection of the sequence number. The field takes one of the following values: nominal sequence - complementary sequence nr 1 - complementary sequence nr 2 - complementary sequence nr 3 - complementary sequence nr 4 - complementary sequence nr 5 - complementary sequence nr 6 - complementary sequence nr 7 -
LDL Type. The field takes one of the following values: normal LDL - mixed LDL